Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
22
Document no
.
M1144956-003
4.
Net Unit and Synchronization Unit
Check that:
•
screws are tightened properly
•
the connector pins are intact
"
5.
Front Unit
Check that:
•
the Connector Board screws are tightened properly and the cable connectors are
clean and intact
•
the Inverter Board screws are tightened properly and the cable connectors are
clean and intact
•
the Backlight cables are intact and attached properly
•
the keyboard foil cables are intact and attached properly to connectors on the
Connector Board
•
the Connector Board foil cable is intact and attached properly to the connector on
the Central Processing Board
•
the video signal foil cable is intact and attached properly to its connector on the
Central Processing Board
Reattach the Front Unit.
"
6.
External parts
Check external parts of the Display Unit:
•
all stickers are intact
•
the LCD display screen is intact
•
the ComWheel cover is intact and attached properly
•
all mechanical parts are functioning properly
Connect the Display Unit to the Frame Unit.
"
7.
Inserting module
Check that the module used goes in smoothly and locks up properly in both module slots.
Leave the module connected.
"
8.
Fan filter
Clean or replace the Display Unit fan filter.
"