Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
32
Document no
.
M1144951-004
Figure 12
Connecting the gas module to the scavenging connector of S/5
Avance
NOTE: Refer to the anesthesia machine’s documentation to find out where and how the sample
gas can be connected.
Connecting directly to the vacuum scavenging system
1.
Connect the exhaust line to the monitor’s sample gas outlet.
2.
Connect the exhaust line only to an open scavenging system where gas is removed at
room pressure.
NOTE: Do not connect the monitor directly to a vacuum scavenging system.
Returning sampling gas to the patient circuit
The sampling gas can also be returned to the patient circuit. If you use the S/5 Anesthesia
Delivery Unit (ADU), you need an optional adapter connected to the patient breathing tubes.
Take special care when returning sample gas to the patient circuit. For further information,
please contact your GE Healthcare distributor.
NOTE: If E-miniC is being used, do not return the sample gas to the patient circuit.