2.1.5 Recorder Unit
1.
Clean the recorder.
−
Open the paper compartment hatch and remove the paper roll, if installed.
−
Remove any paper chaff from the paper compartment.
−
Clean the thermal printhead and the small glass window in front of the static brush
with a cotton swab dipped in isopropyl alcohol, if necessary.
NOTE: Avoid contact with the rubber paper roller. Be careful to limit the application of alcohol to
the thermal printhead and the window.
−
Reinstall the paper roll.
"
2.1.6 Modules with NIBP measurement
1.
Check the NIBP pump filter. Replace the filter, if necessary.
NOTE: M-NIBP Rev. 00-03 do not contain a NIBP pump filter.
Plug the module back into the frame.
"
2.2 Functional inspection
2.2.1 General
1.
.Connect the mains power cord
Check that the LED above the mains switch is lit continuously (full charged batteries) or
periodically (charging batteries).
2.
Switch the monitor on.
Check that the monitor starts up properly, i.e. the alarm LEDs and the display unit fan turn
on shortly, normal start-up sound is heard from the loudspeaker and the monitoring
screen appears.
No error messages should appear on the screen.
3.
Configure the screen for the parameters that are connected.
Check that all the connected modules are recognized, i.e. the required parameter
information is shown on the screen.
If installed, the recorder should record two lines of start-up information.
Preset the measurement settings for those parameters that are connected, for example:
Others -
EEG - Montage - EEG Channels - 4
- Montage type - Bip
Others -
EEG - EEG Setup - Numeric 1 - MF
- Numeric 2 - Ampl.
Others
- EP - Cycle - Cont.
- AEP Setup - AEP Channels - 2
- Responses -100
- Stim. Frequency - 1.1Hz