Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
12
Document no
.
M1144958-001
3.2.3 Before beginning
•
Make sure that the monitor is switched to standby.
•
Disconnect the power cord and press the service reset switch until the LED next to it turns
off.
•
If the monitor is connected to the Datex-Ohmeda Network, disconnect the Mon-Net
cable. Take out any memory cards, if installed.
•
Remove the AC/DC Power Unit.
•
Disassemble the Side Frame Unit.
•
Detach the battery cables from the Power Management Board.
3.2.4 Visual inspection
1.
Side Frame Unit
Check the Side Frame Unit:
•
screws are tightened properly
•
cables are connected properly
•
connectors are clean and intact
"
2.
PMB and the Connector Boards
Check the Frame Unit from inside. Check the PMB and the Connector Boards:
•
there are no loose objects inside the Frame Unit
•
all screws are tightened properly
•
all connectors are clean and intact
•
cables are connected properly
"
3.
Frame Unit reattached
Reconnect the battery cable.
Reattach the Side Frame Unit.
Reattach the Frame Unit back plate.
Reattach the AC/DC power unit.
"
NOTE: Make sure that all insulation plates are placed correctly before reattaching the rear
cover.
4.
Battery slot
Internal parts under bottom cover
Detach the bottom cover and check internal parts:
•
the battery cable is connected properly