Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
26
Document no
.
M1144956-003
22. ComWheel check
Turn the ComWheel clockwise and counter clockwise and check that each step generates
a sound from the loudspeaker and the corresponding values at the bottom of the menu
increase.
Select
Dummy Press.
Press the ComWheel and check that the press generates a sound
and the corresponding value in the menu increases.
"
23. K-ARK connector
Check that the command board section of the Anesthesia Keyboard functions normally
through the K-ARK connector.
"
24. Module communication
Connect a patient simulator to the module.
Enter the ESTP: ECG service menu:
Monitor Setup
Install Service
(Password 16-4-34)
Service
(Password 26-23-8)
Parameters
ESTP: ECG
Check that the Timeouts, Bad checksums and Bad c-s by mod values do not increase
faster than by 5 per second.
Change the module into the other slot. When the module is recognized, check that the
values do not still increase faster than by 5 per second.
"
25. Battery
Enter the
Battery
service menu:
Monitor Setup
Install Service
(Password 16-4-34)
Service
(Password 26-23-8)
Frame
Power Supply
Battery
Check that the information on the page is correct.
"
26. Battery voltage
Check that the shown Battery Voltage value is within 10 – 17V.
"