![GE Datex-Ohmeda S/5 Series Technical Reference Manual Download Page 293](http://html.mh-extra.com/html/ge/datex-ohmeda-s-5-series/datex-ohmeda-s-5-series_technical-reference-manual_128583293.webp)
Frame Unit for F-CM(C)1 (Rev. 05), F-CM(C)REC1 (Rev. 01)
19
Document no. M1144958-001
3.3 Disassembly and reassembly
WARNING
The S/5 Compact Monitor is always energized by an internal battery. A
short circuit may cause internal damage. Do not touch any exposed wiring
or conductive surface inside, as this may cause an electric shock.
See the exploded view of the S/5 Compact Monitor Frame Unit in the ‘Spare Parts’ slot to get
quick unerstaning of the structure.
3.3.1 To remove the batteries
1.
Open the Bottom Plate of the Frame Unit by opening the eight screws.
2.
Detach the battery cables and remove the batteries.
NOTE: Remember the order of the batteries connected to the PMB board.
3.3.2 To remove the AC/DC Unit
1.
Open the four screws holding the unit at the back of the Compact Monitor .
2.
Pull the AC/DC Unit carefully to avoid damaging the connector.
NOTE: It is not allowed to repair the AC/DC power unit in the field.
Return the faulty AC/DC power unit to GE Healthcare for service.
3.3.3 To disassemble the Side Frame Unit
1.
Detach/remove the AC/DC Unit.
2.
Open the three screws from the left side of the Frame Unit back plate.
3.
Loosen the long screw at the upper left corner of the Frame Unit.
4.
Open the two screws at the bottom left side of the Frame Unit.
5.
Pull the Side Panel Unit carefully from the connector to the PMB Board and out of the
frame.
NOTE: When reassembling the Side Panel Unit, be careful not to damage the components on
the PMB Board.
3.3.4 To detach/remove the PMB Board (Power Managemet Board)
NOTE: It is important to remove the batteries before starting to remove the PMB Board!
NOTE: Remember the order of batteries connected to the PMB Board.
1.
Disassemble the Side Panel Unit.
2.
Open the Bottom Plate of the Frame Unit by opening the eight screws and remove the
batteries.
3.
Open the two screws holding the connector to the AC/DC Unit.
4.
Open the two screws holding the connector to the Display Unit.
5.
Detach the Power and Data cables from the PMB Board.
6.
Open the eight screws holding the PMB Board.
7.
Pull the PMB Board carefully to detach it from the Module Mother Board.