![GE Datex-Ohmeda S/5 Series Technical Reference Manual Download Page 276](http://html.mh-extra.com/html/ge/datex-ohmeda-s-5-series/datex-ohmeda-s-5-series_technical-reference-manual_128583276.webp)
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
2
Document no
.
M1144958-001
1
Specifications
Dimensions
335 x 205 x 340 mm /13.2 x 8.1 x 13.4in
(W x D x H)
Weight
8.5 kg (19.4 lbs) (w/o modules)
Electrical Requirements
Power supply
Rated voltage range and frequencies 100 to 240 VAC 50/60 Hz
(limits 90-264 VAC)
Stability
±10 % of nominal voltage
Power consumption
140 VA (max)
Fuses
2x2A, T/SLOW, 250 V
Grounding
Hospital grade
Interruptability
Data memory and alarm settings are saved up to 15 minutes
after monitor standby
Battery
Battery type
NiMH, 3.5Ah
Battery charging time
10 hours to full capacity
Battery operation
up to 1.5 hours typical on full charge at 23 °C / 73 °F
Recharging
when connected to mains power
Battery operaton alarm
early warning
‘Battery low’ note
final warning
‘Battery low’ alarm
Battery state indicator
Capacity bar displayed on the screen
Environmental Requirements
Operating temperature
10 - 35 °C / 50 - 95 °F
Storage temperature
-10 - 50 °C / 14 - 122 °F
Atmospheric pressure
500 - 800 mmHg (660 - 1060 mbar, 66 - 106kPa)
Humidity
10 - 90 % non-condensing
In-built recorder
Power consumption
3 W
Principle
Thermal array
Print resolution
Vertical
8 dots/mm (200 dots/inch)
Horizontal
32 dots/mm (800 dots/inch) at a speed of 25 mm/s and slower
Paper width
50 mm, printing width 48 mm
Traces
Selectable 1, 2, or 3 traces
Print speed
1, 6.25, 12.5, 25 mm/s