![GE Datex-Ohmeda S/5 Series Technical Reference Manual Download Page 287](http://html.mh-extra.com/html/ge/datex-ohmeda-s-5-series/datex-ohmeda-s-5-series_technical-reference-manual_128583287.webp)
Frame Unit for F-CM(C)1 (Rev. 05), F-CM(C)REC1 (Rev. 01)
13
Document no. M1144958-001
•
the batteries are properly in place
•
there are no loose objects inside the battery slot
Reattach the bottom cover.
"
5.
External parts
Check external parts:
•
the four rubber pads under the monitor are all in place
•
all stickers are intact
•
the Display Unit is attached properly
•
all mechanical parts are functioning properly
•
the service reset switch is intact
•
the handle is attached properly
•
the monitor frame is intact
"
6.
Module connections
Check that the module in use goes in smoothly and locks up properly in both module
slots.
"
Leave the module connected.
7.
Fan filter
Clean or replace the Frame Unit Fan filter.
"
8.
Recorder
Clean the recorder unit, if installed:
Open the paper compartment cover and take out the paper roll.
Clean the paper compartment off of possible paper chaff.
Clean the thermal printhead and the small glass window before the static brush with a
cotton swab dipped in isopropyl alcohol. Avoid contact with the rubber paper roller.
NOTE: Be careful to limit the application of alcohol to the thermal printhead and the window.
Attach a new paper roll into the recorder and close the cover.
"
9.
Primary fuses
Disconnect the primary fuse holder. Check that:
•
the fuse compartment, the fuse holder and the fuses are clean and intact