Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
6
Document no
.
M1144955-001
1.2.2 Network Config
The DRI Level
is for setting the monitor's
network communication. The network
communication is set according to the network
software used (e.g. S-CNET01).
Network software S-CNET99 -> DRI Level = 1999
Network software S-CNET01 -> DRI Level = 2001
Network software S-CNET02 -> DRI Level = 2001
or 2002 (WLAN)
Network software L-NET03 -> DRI Level = 2003
Network software L-NET05 -> DRI Level = 2005
NOTE: The DRI Level 1999 is not selectable in all
monitor software versions.
Virtual Plug ID
indicates the monitor’s location
(Network ID). Virtual Plug ID is automatically
defined when downloading the WLAN option to
the monitor.
1.2.3 Ethernet
The
Ethernet Status
view shows the general
status of the ethernet network communication.
Driver
: Ethernet chip name (DP83902,
DP83907, Ethernet Driver LAN911x)
Cable
: Indicates if the ethernet cable is
connected
.
EthernetAddr
:
Monitor’s ethernet address.
Speed
: Indicates the current ethernet
communication speed.
The service data related to the ethernet status
view is described in the following table.