7/48
EN
4.8.
Setting of pressure:
Prior to releasing gas into the system, check the following:
1.
Correct Cylinder Pressure Regulator
2.
All gauge pointers at zero
3.
Pressure adjusting knob (5) fully screwed out (anti-clockwise)
4.
All shut-off valves are closed.
Slowly open the cylinder valve (1), ensure that the cylinder pressure is showing on the content’s gauge (3), open the
downstream (Blowpipe) valve, and with the Pressure adjusting knob (5) rotate clockwise to set the required working
pressure using the outlet gauge (4).
4.9.
Indications to Cylinder pressure regulator having Flowmeter setting of the required flow: Open slowly cylinder
valve (1). The inlet gauge (3) is indicating the cylinder pressure. Open valve (6) and corresponding shut-off valve of the
implement. Set the desired flow by adjustment valve (5).
TERMINATION OF OPERATION
•
Close Cylinder valve (1), unwind (anti-clockwise) the Pressure adjusting knob (5), vent the pressure in the system
by opening the downstream (Blowpipe) valves.
•
Once the system has been purged, close the downstream valves.
OPERATING AND MAINTENANCE INSTRUCTIONS
6.1.
Protect the pressure control regulator from damage (check visually at regulator intervals).
6.2.
The pre-set adjustment of the safety valve (8) must not be changed.
6.3.
Ensure that gaskets, sealing surfaces and gauges are in good condition and free from marks.
6.4.
If malfunction of the regulator (such as increasing outlet pressure at consumption= 0, leaking, defective gauges
or responding safety valve) is suspected put the pressure control regulator out of operation and close the cylinder
valve (1) immediately.
Do not under any circumstances undertake, or allow any repairs by unauthorised personnel.
REPAIRS
7.1.
Repairs of Cylinder Pressure Regulators shall be carried out by competent personnel at an authorised Service or
Repair Workshop. Only genuine spares may be used.
7.2.
Repairs or modifications carried out by the user or a non-authorised third party will entail loss of liability.
CYLINDER PRESSURE REGULATORS WITH FLOWMETER GAUGE:
Paragraphs 1 to 7 of these Operating Instructions applies also to Cylinder Pressure Regulators with Flow gauge. The
flow set by the Pressure adjusting knob (5) may be read at the Flowmeter gauge (4). Cylinder Pressure Regulator with
flow gauge have an integrated throttle. If the apparatus or equipment to be used downstream also has a Flowmeter
device, please remove it as it will not be calibrated with the Cylinder Pressure Regulator.
WARRANTY
The Standard Warranty period is two years from date of receipt by the Customer (or if this is not known 2 years from
time of the product manufacture shown on the product). The standard warranty is only valid for products handled
according to Instruction for use (IFU) and general industry good practice and standards.
5.
6.
7.
8.
9.
NOTE: PICTURES IN THIS INSTRUCTION FOR USE ARE ILLUSTRATIVE ONLY. THEY MAY DIFFER FROM THE REAL LOOK OF YOUR
PRODUCT!
Summary of Contents for DINCONTROL
Page 2: ...FIG 1...
Page 4: ...FIG 1...
Page 43: ...43 48 CN 1 5 6 1 6 2 8 6 3 6 4 1 7 1 7 2 1 7 5 4 5 6 7 8 9...
Page 46: ......
Page 47: ......
Page 48: ...Doc Nr IFU0033 Date of issue 2016 08 30 Rev 00 TI A5 CB V1...