Remote Control - Description of Commands
Commands and Queries
4
4-51
Invalidated by
-
[SENSe:]SWEep1|2:FUNCtion <function>{,<function>}
Description
Specifies function list for memory recording. The maximum number of functions is 20. If
the list exceeds device capability, the sample factor is automatically increased.
For SWEep1 (REALtime) function list only sampled values can be specified.
For SWEep2 (AVERage) function list any functions from the standard instrument
function list can be specified.
The function list is stored in a saved configuration and is reloaded at PowerOn or with
*RCL.
Parameter(s)
<function>
SWEep1 (REALtime) valid functions:
VOLTage1..6[:DC]
CURRent1..6[:DC]
POWer1..6[:ACTive]
TORQue[1..4]
SPEed[1..4]
POWer[1..4]:MECHanical
SWEep2 (AVERage) valid functions:
Any functions from the standard instrument function list can be specified.
*RST state
“VOLTage1”
Example(s)
SWE1:FUNC “VOLT1”, VOLT2”,”VOLT3”
Invalidates
-
Invalidated by
-
SENSe2 Subsystem (Option Process Interface)
The SENSe2 subsystem controls the settings of the optional Process Interface inputs.
Numeric suffixes of the TORQue|SPEed|POLepairs|TYPe|REFerence nodes correspond
to the index of the 4 motors/generators supported. Channels of the INPut subsystem are
combined into drive index of the SENSe2:xxx subsystem as shown in the table below.
Summary of Contents for NORMA 4000 Operators
Page 6: ...Remote Control Users Guide iv ...
Page 8: ...Remote Control Users Guide vi ...
Page 10: ...Remote Control Users Guide viii ...
Page 26: ...Remote Control Users Guide 2 2 ...
Page 38: ...Remote Control Users Guide 3 2 ...
Page 50: ...Remote Control Users Guide 4 2 ...
Page 150: ...Remote Control Users Guide 4 102 ...
Page 152: ...Remote Control Users Guide 5 2 ...
Page 160: ...Remote Control Users Guide 6 2 ...
Page 184: ...Remote Control Users Guide 6 26 ...