Remote Control
Users Guide
4-40
Example(s)
VOLT3:RANG 25.0
CURR2:RANG? Response:
3.0
*RST state
After reset autoranging is ON, so no fixed range is set.
Invalidates
[SENSe:]CURRent[1..6]|VOLTage[1..6]:RANGe:AUTO ON
Invalidated by
[SENSe:]CURRent[1..6]|VOLTage[1..6]:RANGe:AUTO ON
Averaging
[SENSe:]CURRent[1..6]|VOLTage[1..6]:AC[|:DC]:RANGe:[:UPPer]:LIST?
Description
Queries a list of valid voltage/current ranges available on channel specified by the phase
suffix. The returned list contains the actual ranges of the instrument’s input channels.
Channel scaling factors and external shunts (INPut:GAIN) are not included. For current
channels the list is dependent on the actual setting of INPut:SHUNt (EXTernal or
INTernal).
Response
<range_list>
Comma separated range value list.
Example(s)
CURR2:RANG:LIST?
0.03, 0.1, 0.3, 1, 3, 10
(for setting INPut:SHUNt INTernal)
*RST state
Invalidates
-
Invalidated by
-
[SENSe:]CURRent[1..6]|VOLTage[1..6]:AC[|:DC]:RANGe[:UPPer]:AUTO
ON|OFF|ONCE
Description
Controls the voltage/current autoranging.
Parameter(s)
ON
Autoranging is permanently ON. Monitoring of the STATus:OPERation register will
detect that the range is changed.
OFF
Autoranging is permanently OFF.
ONCE
ONCE is currently unimplemented
Summary of Contents for NORMA 4000 Operators
Page 6: ...Remote Control Users Guide iv ...
Page 8: ...Remote Control Users Guide vi ...
Page 10: ...Remote Control Users Guide viii ...
Page 26: ...Remote Control Users Guide 2 2 ...
Page 38: ...Remote Control Users Guide 3 2 ...
Page 50: ...Remote Control Users Guide 4 2 ...
Page 150: ...Remote Control Users Guide 4 102 ...
Page 152: ...Remote Control Users Guide 5 2 ...
Page 160: ...Remote Control Users Guide 6 2 ...
Page 184: ...Remote Control Users Guide 6 26 ...