Remote Control
Users Guide
4-76
TRIGger:STARt:SOURce BUS | TIME | IMMediate | MANual | SYNC | <function>
Description
Specifies the start trigger source.
Parameter(s)
BUS
Will trigger when a *TRG command is received
TIME
Will trigger at exact time
IMMediate
No waiting for an event occurs.
MANual
The signal is user-generated by pressing the front panel “MEM" key.
SYNC
Trigger occurs whenever an edge of the synchronization signal is detected. This source
is valid only for REALtime sweep (to use the EXTernal signal jack as a trigger the
SYNC source must be set to EXTernal).
<function>
Condition on an averaged measurement function causes the trigger.
*RST state
IMM
Example(s)
TRIG:STAR:SOUR IMM
TRIG:STAR:SOUR? Response:
IMM
Invalidates
-
Invalidated by
-
TRIGger:STARt:TIME <yyyy,MM,dd,hh,mm,ss>
Description
The memory recording will start when the instrument’s internal time reaches the
specified value.
Parameter(s)
yyyy Year
MM Month
dd Day
hh
Hours in 24 hour notation
mm Minutes
ss
Seconds (integer value)
*RST state
1970,1,1,0,0,0
Example(s)
TRIG:STAR:TIME 2002,01,01,11,00,00
TRIG:STAR:TIME? Response:
2002,01,01,11,00,00
Invalidates
-
Invalidated by
-
Summary of Contents for NORMA 4000 Operators
Page 6: ...Remote Control Users Guide iv ...
Page 8: ...Remote Control Users Guide vi ...
Page 10: ...Remote Control Users Guide viii ...
Page 26: ...Remote Control Users Guide 2 2 ...
Page 38: ...Remote Control Users Guide 3 2 ...
Page 50: ...Remote Control Users Guide 4 2 ...
Page 150: ...Remote Control Users Guide 4 102 ...
Page 152: ...Remote Control Users Guide 5 2 ...
Page 160: ...Remote Control Users Guide 6 2 ...
Page 184: ...Remote Control Users Guide 6 26 ...