Remote Control - Description of Commands
Commands and Queries
4
4-19
Parameter(s)
-
*RST state
-
Example(s)
CALC:INT:CLE
Invalidates
-
Invalidated by
-
CALCulate:INTegral:CLEar:AUTO ON | OFF
Description
This command controls automatic zeroing of the integrated functions are set to zero.
ON
Integrated values get cleared at the start of the integration. All values get set to zero at
the start of integration.
OFF
Automatic zeroing of integrated values is disabled.
Example(s)
CALC:INT:CLE:AUTO ON
CALC:INT:CLE:AUTO? Response:
ON
*RST state
ON
Invalidates
-
Invalidated by
-
CALCulate:INTegral:STARt:SOURce CMD | TIME | MAN
Description
Specifies integration start condition.
Parameter(s)
CMD
Integration starts upon receipt of the CALCulate:INTegral:STARt[:IMMediate] command.
TIME
Integration will start at a time specified by command CALCulate:INTegral:STARt:TIME
MAN
Integration will start when user presses F1 key on the front panel from within integration
measurement screen of the instrument
Example(s)
CALC:INT:STAR:SOUR CMD
CALC:INT:STAR:SOUR? Response:
CMD