Remote Control
Users Guide
4-24
Example(s)
CALC:POW:EFF:REF “POW460”, “POW1”
CALC:POW:EFF:REF? Response: “POW460”, “POW1”
Invalidates
-
Invalidated by
-
CALCulate:POWer:CORRected STAR | DELTa
Description
Selects phase-to-neutral or phase-to-phase voltages for the calculation of no load loss
measurements on transformers according to IEC60076-1 (measurement function
POWer[1|2|3|4|5|6|460]:CORRected).
In a 3-phase instrument, STAR is not possible together with W2 (Aron), parameter will
be set to DELTa automatically. If there are more than 3 phases, STAR may still be
selected in mode W2, but only applies to phase 4 and up. Pcorr from the first system is
not available then.
This command is only accepted by firmware version V1.4 and up.
Parameter(s)
STAR
Use phase-to-neutral voltages (wye connection scheme)
DELTa
Use phase-to-phase voltages (delta connection scheme)
*RST state
CALCulate:POWer:CORRected?: STAR
Example(s)
CALC:POW:CORR DELT
CALC:POW:CORR? Response:
DELT
Invalidates
-
Invalidated by
ROUTe:SYST "2W" (3-phase instruments only)
DISPlay Subsystem
The DISPlay subsystem contains commands to control the display.
Command Parameter
Default
Value/Unit
Remark
:DISPlay
[:WINDow]
[:STATe]
:USER
: FUNCtion
ON | OFF
<function list>
OFF
Summary of Contents for NORMA 4000 Operators
Page 6: ...Remote Control Users Guide iv ...
Page 8: ...Remote Control Users Guide vi ...
Page 10: ...Remote Control Users Guide viii ...
Page 26: ...Remote Control Users Guide 2 2 ...
Page 38: ...Remote Control Users Guide 3 2 ...
Page 50: ...Remote Control Users Guide 4 2 ...
Page 150: ...Remote Control Users Guide 4 102 ...
Page 152: ...Remote Control Users Guide 5 2 ...
Page 160: ...Remote Control Users Guide 6 2 ...
Page 184: ...Remote Control Users Guide 6 26 ...