Remote Control
Users Guide
4-60
Example(s)
SENS2:TYP1 MOT
SENS2:TYP3? Response:
GEN
*RST state
MOT
Invalidates
-
Invalidated by
-
SENSe2:POLepairs[1..4] <value>
Description
Specifies the number of polepairs of the drive. This setting is used for slip calculation.
Parameter(s)
<value>
Valid Range: 1 to 999
Example(s)
SENS2:POL3 2
SENS2:POL1? Response:
1
*RST state
1
Invalidates
-
Invalidated by
-
SENSe2:REFerence[1..4][:POWer] <function>
Description
Specifies the measured electrical power used for efficiency calculation.
Parameter(s)
<function>
can be any of the averaged active powers measured by the instrument:
"POWer[1..6|460][:ACTive]"
Example(s)
SENS2:REF3 "POW1"
SENS2:REF2? Response:
"POW"
*RST state
"POW"