Remote Control
Users Guide
4-72
Response
When FORMat:TRANspose is ON, values are grouped by functions:
<interval1>,<interval2>,<interval3>,..............<interval1>,<interval2>,<interval3>,….
func1
func2
When
FORMat:TRANspose
is
OFF
, values are grouped by intervals:
<func1>,<func2>,<func3>,.........<func1>,<func2>,<func3>,….
interval1
interval2
*RST state
There is no response to this command after reset.
Example(s)
TRAC:DATA? Response:
1.2345E+01,2.3456E+01,….
Invalidates
-
Invalidated by
-
TRACe[:DATA]:STATus? [<block>[,<count>[,<offset>[,<sparsing>]]]]
Description
Reads data and status from memory.
Data are returned first, then followed by the status information.
The measurement status information indicates the validity of the measurement. The status
information is appended to the set of measured values. Number of status values is equal
to the number of returned measured values. The format of the status information is
controlled by the FORMat:STATus commands.
For best performance, use binary output with 32-bit wide floating-point numbers
(FORMat[:DATA] REAL,32), normal (instrument’s native) data byte ordering
(FORMat:BORDer NORMal) and 8-bit wide integer status numbers
(FORMat[:DATA]:STATus INT,8).
Parameter(s)
See TRACe[:DATA]?.
Status Values
The returned status values are integers. The measurement status values are appended to
the end of the measurement results. The status value is bit mask integer and can be a
Summary of Contents for NORMA 4000 Operators
Page 6: ...Remote Control Users Guide iv ...
Page 8: ...Remote Control Users Guide vi ...
Page 10: ...Remote Control Users Guide viii ...
Page 26: ...Remote Control Users Guide 2 2 ...
Page 38: ...Remote Control Users Guide 3 2 ...
Page 50: ...Remote Control Users Guide 4 2 ...
Page 150: ...Remote Control Users Guide 4 102 ...
Page 152: ...Remote Control Users Guide 5 2 ...
Page 160: ...Remote Control Users Guide 6 2 ...
Page 184: ...Remote Control Users Guide 6 26 ...