Remote Control - Description of Commands
Commands and Queries
4
4-37
Invalidates
-
Invalidated by
-
SENSe Subsystem
The SENSe subsystem controls the averaging capability of the instrument and calculation
of the basic averaged values. Numeric suffixes of the VOLTage|CURRent nodes
correspond to electrical phases. Channels of the INPut subsystem are combined into
phases of the SENSe:VOLTage|CURRent subsystem as shown in the table below.
If the channel suffix is omitted, the command applies to phase 1.
Input Channel
Suffix
Electrical Phase
Sense Subsystem Phase Suffix
INPut1 (current)
INPut2 (voltage)
L1 SENSe:VOLTage|CURRent1
INPut3 (current)
INPut4 (voltage
L2 SENSe:VOLTage|CURRent2
INPut5 (current)
INPut6 (voltage)
L3 SENSe:VOLTage|CURRent3
INPut7 (current)
INPut8 (voltage)
L4 SENSe:VOLTage|CURRent4
INPut9 (current)
INPut10 (voltage)
L5 SENSe:VOLTage|CURRent5
INPut11 (current)
INPut12 (voltage)
L6 SENSe:VOLTage|CURRent6
The SENSe node is the default node of the root level of the command tree. The default
node of the SENSe subsystem is POWer. All SENSe phase-related settings are common
to both AC and DC couplings, so the AC | DC node can be omitted assuming that DC is
used. The :AC[|:DC] node is implemented for compatibility reasons only.
Summary of Contents for NORMA 4000 Operators
Page 6: ...Remote Control Users Guide iv ...
Page 8: ...Remote Control Users Guide vi ...
Page 10: ...Remote Control Users Guide viii ...
Page 26: ...Remote Control Users Guide 2 2 ...
Page 38: ...Remote Control Users Guide 3 2 ...
Page 50: ...Remote Control Users Guide 4 2 ...
Page 150: ...Remote Control Users Guide 4 102 ...
Page 152: ...Remote Control Users Guide 5 2 ...
Page 160: ...Remote Control Users Guide 6 2 ...
Page 184: ...Remote Control Users Guide 6 26 ...