Remote Control - Description of Commands
Commands and Queries
4
4-69
*RST state
No reset condition
Invalidates
TIM:RES:TIME?
Invalidated by
-
TIMer:RESet:AUTO ON | OFF
Description
Controls whether the instrument's internal timer is reset automatically when ARMed. In
order to maintain absolute timebase for sequenced memory measurements,
TIMer:RESet:AUTO must be set to OFF, so that subsequent
INITiate[:IMMediate]:NAME:STARt commands do not reset the timer.
<currently unimplemented>
Parameter(s)
ON INITiate[:IMMediate]:NAME:STARt resets the timer.
OFF INITiate[:IMMediate]:NAME:STARt does not reset the timer.
Example(s)
TIM:RES:AUTO ON
TIM:RES:AUTO? Response:
1
*RST state
ON
Invalidates
TIM:RESet:TIME?
Invalidated by
-
TIMer:RESet:TIME?
Description
Queries the absolute time of the last timer reset.
Reponse
<year>,<month>,<day>,<hours>,<minutes>,<seconds>
Year is in four-digit numberic format. Hours are in 24 hour notation.
*RST state
Has no reset value
Example(s)
TIM:RES:TIME?
Invalidates
-
Invalidated by
-