Remote Control
Users Guide
4-62
SOURce:VOLTage[1..4][:LEVel][:IMMediate][:AMPLitude] <value>
Description
Selects the output voltage for FIXed mode.
Parameter(s)
<value>
Valid Range: -10.3 to 10.3 V
*RST state
0.0
Example(s)
SOUR:VOLT4 5.3
SOUR:VOLT1? Response:
-2.5
Invalidates
-
Invalidated by
-
SOURce:VOLTage[1..4]:FEED <function>
Description
Specifies the output reference function for VARiable mode.
Parameter(s)
<function>
Any valid averaged measurement function of the instrument.
*RST state
"VOLTage1"
Example(s)
SOUR:VOLT2:FEED "POW2:APP"
SOUR:VOLT4:FEED? Response: "CURR3:MEAN"
Invalidates
-
Invalidated by
-
Scaling
SOURce:VOLTage[1..4]:GAIN <value>
Description
Specifies the scaling for the output. The difference between the actual value of the
reference function and the ZERO value is multiplied by this factor to calculate the output
voltage.