Remote Control - Description of Commands
Commands and Queries
4
4-35
Parameter(s)
INTernal
Internal shunts up to 10A are used.
EXTernal
External shunt is connected. Shunt factor must be specified with command INPut:GAIN.
*RST state
INTernal for all channels
Example(s)
INP1:SHUN EXT
INP3:SHUNt? Response:
INT
Invalidates
-
Invalidated by
-
INPut[21..28]:TYPe VOLTage | FREQuency (Option Process Interface)
Description
This command sets the input type to match the sensor type.
Parameter(s)
VOLTage
DC signal in range of +/- 10 V is expected on the input.
FREQency
AC signal with frequency in range 1 Hz to 200 kHz is expected on the input.
*RST state
FREQuency
Example(s)
INP21:TYP FREQ
INP21:TYP? Response:
FREQ
Invalidates
-
Invalidated by
-
OUTPut Subsystem
The OUTPut subsystem controls the characteristic of the SYNC output.
Command Parameter
Default
Value/Unit
Remark
:OUTPut9
[:STATe]
ON | OFF
OFF
SYNC output