Remote Control - Description of Commands
Commands and Queries
4
4-13
CALCulate:TRANsform:FREQuency[:STATe] ONCE
Description
Initiates a single calculation of the frequency transform (spectrum), i.e. the instrument
calculates spectrum only upon receipt of this command. An attempt to perform the
spectrum calculation while SENSe:SWEep1[:STATe] is ON (memory storage of samples)
will generate error "-221, Settings conflict".
Parameter(s)
ONCE
Example(s)
CALC:TRAN:FREQ ONCE
*RST state
-
Invalidates
-
Invalidated by
-
CALCulate:TRANsform:FREQuency:MODE FFT | DFT
Description
Selects the harmonics calculation method.
Parameter(s)
FFT
Calculates an FFT amplitude spectrum. Number of lines is determined by the
instrument and depends on the stop frequency, anti-aliasing filters setting and
instrument model.
DFT
Calculates a DFT amplitude spectrum, i.e. the amplitude of the calculated
fundamental frequency and its integer multiples. Number of lines (harmonics)
is always 41 including the DC component.
Example(s)
CALC:TRAN:FREQ:MODE FFT
CALC:TRAN:FREQ:MODE? Response:
FFT
*RST state
FFT
Invalidates
-
Invalidated by
-
Summary of Contents for NORMA 4000 Operators
Page 6: ...Remote Control Users Guide iv ...
Page 8: ...Remote Control Users Guide vi ...
Page 10: ...Remote Control Users Guide viii ...
Page 26: ...Remote Control Users Guide 2 2 ...
Page 38: ...Remote Control Users Guide 3 2 ...
Page 50: ...Remote Control Users Guide 4 2 ...
Page 150: ...Remote Control Users Guide 4 102 ...
Page 152: ...Remote Control Users Guide 5 2 ...
Page 160: ...Remote Control Users Guide 6 2 ...
Page 184: ...Remote Control Users Guide 6 26 ...