Remote Control
Users Guide
4-48
[SENSe:]SWEep1|2:POINTS?
Description
Before recording is initiated or finished, this command returns max. number of points per
function that will be recorded. For synchronized SWEep2 (AVERage) recordings, this
value is calculated as:
Configured recording time / nominal averaging interval / sample factor
( SWEep2:TIME? / APER? / SWEep2:SFACtor? )
The actual max. number of points that will be recorded depends on the variations of
frequency of the measured signal.
Once the recording is finished, this command returns actual number of points per
function recorded.
Response
<count>
*RST state
Maximum available number of points. Depends on amount of available memory on the
instrument.
Example(s)
SWE1:POINTS?
Invalidates
-
Invalidated by
-
[SENSe:]SWEep1|2:OFFSet:TIME <value> | MAX
Description
This command specifies pretrigger length in seconds.
Parameter(s)
<value> | MAX
Pretrigger length must be greater or equal to zero and smaller than the record length
specified with [SENSe:]SWEep1|2:TIME
*RST state
0.0
Example(s)
SWE1:OFFS:TIME 1.0
Invalidates
-
Invalidated by
-
Summary of Contents for NORMA 4000 Operators
Page 6: ...Remote Control Users Guide iv ...
Page 8: ...Remote Control Users Guide vi ...
Page 10: ...Remote Control Users Guide viii ...
Page 26: ...Remote Control Users Guide 2 2 ...
Page 38: ...Remote Control Users Guide 3 2 ...
Page 50: ...Remote Control Users Guide 4 2 ...
Page 150: ...Remote Control Users Guide 4 102 ...
Page 152: ...Remote Control Users Guide 5 2 ...
Page 160: ...Remote Control Users Guide 6 2 ...
Page 184: ...Remote Control Users Guide 6 26 ...