Remote Control
Users Guide
4-88
Parameter(s)
bits 0 to 5
Voltage channels (1, 3, 5, 7, 9, 11) enable overload event.
bits 8 to 13
Voltage channels (1, 3, 5, 7, 9, 11) enable underload event.
*RST state
0
Example(s)
STAT:QUES:VOLT:ENAB 16191
STAT:QUES:VOLT:ENAB? Response:
16191
Invalidates
-
Invalidated by
-
STATus:QUEStionable:CURRent:CONDition?
Description
Returns the contents of the condition register associated with the status structure defined
in the command. Reading the condition register is nondestructive. The response is (NR1
NUMERIC RESPONSE DATA) (range: 0 through 32767).
Response
<value> is a 16 bit integer number in decimal notation.
bits 0 to 5
Current channels (0, 2, 4, 6, 8, 10) overload.
bits 8 to 13
Current channels (0, 2, 4, 6, 8, 10) underload.
*RST state
Has no effect.
Example(s)
STAT:QUES:CURR:COND? Response: 2 (current overload on phase 2)
Invalidates
-
Invalidated by
-
STATus:QUEStionable:CURRent:PTRansition <value>
Description
Sets the positive transition filter. Setting a bit in the positive transition filter shall cause a
0 to 1 transition in the corresponding bit of the associated condition register to cause a 1
to be written in the associated bit of the corresponding event register. The command
accepts parameter values of either format in the range 0 through 65535 (decimal) without
error. The query response format is <NR1>.
Summary of Contents for NORMA 4000 Operators
Page 6: ...Remote Control Users Guide iv ...
Page 8: ...Remote Control Users Guide vi ...
Page 10: ...Remote Control Users Guide viii ...
Page 26: ...Remote Control Users Guide 2 2 ...
Page 38: ...Remote Control Users Guide 3 2 ...
Page 50: ...Remote Control Users Guide 4 2 ...
Page 150: ...Remote Control Users Guide 4 102 ...
Page 152: ...Remote Control Users Guide 5 2 ...
Page 160: ...Remote Control Users Guide 6 2 ...
Page 184: ...Remote Control Users Guide 6 26 ...