Error Codes
12-4 Peltier Low Temp. Incubators
Fisher Scientific
Sensor Error (E113)
Sample sensor damaged.
The actual measured value
is out of range.
Audible alarm activated,
alarm relay energized,
message shown on dis-
play.
Call service.
ADC Error (E114)
A/D converter does not
supply a plausible output.
Measurement across refer-
ence resistor has failed.
ADC converter may be
defective.
Audible alarm activated,
alarm relay energized,
message shown on dis-
play. All control circuits dis-
abled.
Call Service.
Watchdog error (E115)
Watchdog test failed on
power-up.
Audible alarm activated,
alarm relay energized,
message shown on dis-
play.
Call service.
*
Clearing should mute the audible alarm, de-energize the alarm relay, and clear the message from the control panel display.
Table 12-1 Fisher Scientific low temperature incubator Error Codes
Error Message & Code
Root Cause
Alarm Response
Alarm Clearing Instructions
*