Technical Data
Fisher Scientific
Peltier Low Temp. Incubators 13-3
IP protection system
IP 20
Protection class
I
Overvoltage category according
to IEC 60364-4-443
II
Device fusing building site
A
16
Device fusing on PCB
A
2x15
Environmental conditions
Ambient temperature range
°C (°F)
18-28 (64-82)
Max. rel. humidity in service at
or below 28°C (82°F), non
condensing
% r.H.
70
Storage temperature range
°C (°F)
20-60 (68-140)
Max. humidity in storage, non
condensing
% r.H.
90
Post-transport acclimation time
h
2
Noise level, measuring point:
device-distance 1.0 m, height
1.6m)
dB(A)
< 50
Pollution degree according to
IEC EN 61010-1
2
Site conditions
Maximum altitude above sea
level
m/y NN
2000/2187
Minimum side clearance
mm/in
150/5.9
150/5.9
Minimum front clearance
mm/in
814/32
810/31.9
Minimum back wall clearance
mm/in
130/5.1
150/5.9
Minimum bottom clearance
mm/in
200/8
-/-
Minimum top clearance
mm/in
300/12
360/14.2
Models
Isotemp 178L
Isotemp 381L