Handling and Control
Settings
Fisher Scientific
Peltier Low Temp. Incubators 7-9
Settings
The
Settings
menu item opens a sub-menu populated with various commands for viewing
general status information on the Fisher Scientific unit and setting for the operation of the
incubator or its display window:
• Read access to error log
• Calibrating the incubator
• Toggling the temperature display unit between °C and °F
• Entering a configuration control code
Instructions for using these features are given in the following.
Error Log
Users calling customer service for support may be asked by the Fisher Scientific agent to
supply information from the error log of the incubator. It enables the user to browse through
the most recent 22 alarm messages that were caused by hardware or control loop errors. Each
error is displayed with an internal error code.
Error codes and instructions for clearing alarm conditions appear in the section
Table 7-9 Reading the Error Log
Press
to activate the menu bar, then use
or
to select the
Settings
icon and press
to confirm.
The word
Err
appears in the display pane to indicate that the
error log has been selected.
Use the
button to select the first entry in the error log,
numbered
E01
(Error 01).
After a few seconds, the display pane automatically switches
to the internal error code - for example,
109
.
E01 shows the latest fault, E22 shows the oldest fault.
Summary of Contents for Isotemp 178L 120V
Page 1: ......
Page 6: ...Contents iv Peltier Low Temp Incubators Fisher Scientific...
Page 8: ...List of Figures ii Peltier Low Temp Incubators Fisher Scientific...
Page 84: ...Technical Data 13 4 Peltier Low Temp Incubators Fisher Scientific...
Page 88: ...Device Log 15 2 Peltier Low Temp Incubators Fisher Scientific...
Page 91: ...Contact Fisher Scientific Peltier Low Temp Incubators 16 3...