Product Description
Sensing and Control System
4-8 Peltier Low Temp. Incubators
Fisher Scientific
Figure 4-7 PT100 Sensor Isotemp 178L
Figure 4-8 PT100 Sensor Isotemp 378L
The work space temperature sensor provides the inputs to the low temperature incubator’s
built-in controller, which continuously compares the measured values to the user-specified set
value and adjusts the heaters according to the result.
Do not remove the protection hose from the sensor.
Protect the sensor from mechanical damage.
The unit features a thermal protection function that is factory-preprogrammed and not
adjustable. It protects the cultures in the work space from overheating: Thermal protection
kicks in on a brief violation of the upper limit, based on the defined setpoint temperature, at
between 2 and 3 °C (35.6 °F and 37.4 °F) (37 °C (99 °F): 2 °C (35.6 °F), > 50 °C (122 °F):
3 °C (37.4 °F)), automatically reducing the work space temperature to the user-specified set
value and allowing the incubation process to be continued even in case of a controller
malfunction. If the thermal protection is activated, the error message (E111) “Temperature
too high” appears in the display window and an audible alarm is sounded.
When the user acknowledges the error message, the red alarm icon (D4 in
) is illuminated and the Temperature Set Value icon (see
highlighted by a red border to indicate that thermal protection has kicked in.
Summary of Contents for Isotemp 178L 120V
Page 1: ......
Page 6: ...Contents iv Peltier Low Temp Incubators Fisher Scientific...
Page 8: ...List of Figures ii Peltier Low Temp Incubators Fisher Scientific...
Page 84: ...Technical Data 13 4 Peltier Low Temp Incubators Fisher Scientific...
Page 88: ...Device Log 15 2 Peltier Low Temp Incubators Fisher Scientific...
Page 91: ...Contact Fisher Scientific Peltier Low Temp Incubators 16 3...