Cleaning and Disinfection
Wipe / Spray Disinfection
Fisher Scientific
Peltier Low Temp. Incubators 9-3
Preparing the Manual Wipe/Spray Disinfection
Health hazard
The surfaces of the work space may be contaminated. Contact
with contaminated cleaning liquids may cause infections.
Disinfectants may contain harmful substances.
When cleaning and disinfecting, always observe the safety
instructions and hygiene guidelines!
Predisinfection
1. Remove all samples from the work space and store them in a safe place.
2. Spray disinfectant onto the surfaces of the work space and of the accessories or wipe the surfaces
clean using disinfectant.
Risk of injury caused by breaking of
glass panel
The glass panel may only be removed by properly trained and
authorized specialists.
The glass panel must be held by two persons.
3. Allow time for the disinfectant to act as specified by the manufacturer.
Disinfecting hard-to-reach components
Spray the sensor and other hard-to-reach components with
disinfectant!
Check to ensure that the air baffles are securely screwed into place
after cleaning and moving the unit.
• Wear safety gloves.
• Wear safety goggles.
• Wear mouth and respiratory system protection gear to protect your
mucous membranes.
• Observe the safety instructions of the disinfectant’s manufacturer
and the hygiene supervisor.
Summary of Contents for Isotemp 178L 120V
Page 1: ......
Page 6: ...Contents iv Peltier Low Temp Incubators Fisher Scientific...
Page 8: ...List of Figures ii Peltier Low Temp Incubators Fisher Scientific...
Page 84: ...Technical Data 13 4 Peltier Low Temp Incubators Fisher Scientific...
Page 88: ...Device Log 15 2 Peltier Low Temp Incubators Fisher Scientific...
Page 91: ...Contact Fisher Scientific Peltier Low Temp Incubators 16 3...