Installation
Installing the Anti-tilt Anchor
3-8 Peltier Low Temp. Incubators
Fisher Scientific
Installing the Anti-tilt Anchor
Table-top low temperature incubators
The anti-tilt anchor secures the top device in a stack to a solid part of a building. The anti-tilt
anchor is to be mounted on the side opposite of the door hinges.
Bend the fixing tabs of the anti-tilt anchor up on one side and down on the other by an angle
of approx. 90°.
1. Do not use this position if the door is hinged on this side. Right-hand hinges represent the
standard configuration.
2. Preferred position.
3. Alternative position. Do not use if the door is hinged on this side.
Remove the bracket screws. Use the preferred position, if possible.
Fix the anti-tilt anchor with the bracket side down to the unit.
Position the unit with the anti-tilt anchor to in an angle of approx. 90° +/- 20%.
Take care that the stacking feet of the unit are still in correct place on the lower unit or on the
stacking adapter.
Fix the anti-tilt anchor to a solid part of the building.
Floor stand low temperature incubators
Fisher Scientific floor stand low temperature incubators must always be
attached to the wall using two (2) retaining brackets on the outer left and right
side on the back of the unit.
Summary of Contents for Isotemp 178L 120V
Page 1: ......
Page 6: ...Contents iv Peltier Low Temp Incubators Fisher Scientific...
Page 8: ...List of Figures ii Peltier Low Temp Incubators Fisher Scientific...
Page 84: ...Technical Data 13 4 Peltier Low Temp Incubators Fisher Scientific...
Page 88: ...Device Log 15 2 Peltier Low Temp Incubators Fisher Scientific...
Page 91: ...Contact Fisher Scientific Peltier Low Temp Incubators 16 3...