Keep the power outlet accessible!
To allow a rapid disconnection of power in case of an emergency, make
sure that power outlets remain freely accessible at all times!
Installation procedures
Connecting the RS-232 Interface
5-8 Peltier Low Temp. Incubators
Fisher Scientific
Figure 5-9 AC Power Supply Socket
Condensation
When taking the low temperature incubator into operation for the first
time allow some time before switching on for stabilization to avoid
condensation forming on live parts.
If a high-voltage test is to be performed on the unit, it must first be
heated for around 30 minutes at 50°C.
Connecting the RS-232 Interface
Replacement should only be carried out by skilled and authorized
qualified personnel of electrotechnology/signal engineering!
The RS-232 data communication interface supports the querying of status information and
temperature data from the low temperature incubator by entering basic commands in a
standard terminal window provided by your computer’s operating system. The
interconnection requires a standard RS-232 cable with 9-pin connectors and a straight “1:1”
pinout without any crossed wires, which is not supplied with the low temperature incubator.