Installation
Installing the Anti-tilt Anchor
Fisher Scientific
Peltier Low Temp. Incubators 3-9
Remove the screws.
Attach the end of the retaining bracket that is facing downward to the unit.
Align the device at roughly 90°, +/-20° to the retaining bracket.
Affix the retaining bracket to the wall.
Unsafe part of the building!
Install the anti-tilt anchor to a solid part of the building, which is able for shoring
loads.
The installation has to be carried out by qualified personnel only.
The connection to the building must be carried out with appropriate screws and
dowels according to the consistence of the building part.
Additionally, the following caution notes must be heeded at all times
:
Risk of overheating with stacked devices
To avoid the risk of electrical components and the outer enclosure overheating
or temperature control failing due to insufficient ventilation, do not exceed the
specified stacking height!
Risk of tipping and dropping of stacked devices
You should be aware at all times that stacked devices do not form a stable unit,
even when the stacking pads and frames are correctly used. The top device may
tip over and drop down when being transported in a stack. To avoid injury to
persons and damage to equipment, do not attempt to move stacked devices as
a unit! Separate and move each device one by one, then re-stack them.
Thermo Scientific don‘t give warranty for stacking equipment of other manufac-
turer‘s, this occurs at own risk.
Summary of Contents for Isotemp 178L 120V
Page 1: ......
Page 6: ...Contents iv Peltier Low Temp Incubators Fisher Scientific...
Page 8: ...List of Figures ii Peltier Low Temp Incubators Fisher Scientific...
Page 84: ...Technical Data 13 4 Peltier Low Temp Incubators Fisher Scientific...
Page 88: ...Device Log 15 2 Peltier Low Temp Incubators Fisher Scientific...
Page 91: ...Contact Fisher Scientific Peltier Low Temp Incubators 16 3...