Handling and Control
Settings
7-10 Peltier Low Temp. Incubators
Fisher Scientific
Calibration
The
Settings -> Calibration
menu item enables the user to initiate a temperature calibration
process (see
“Temperature Calibration Procedure”
) for the built-in temperature
sensors and choose whether calibration should be accomplished manually or automatically:
• The
Manual
option allows for entering an absolute temperature directly, as measured - for
example, using an external reference sensor.
Calibration Prerequisites
Maintain the ambient conditions within the specified limits of the
incubator before launching calibration.
Varying ambient conditions may impact the result of the
calibration routine, which may lead to mis adjustment of the
controller and unreliable temperature control operation.
Press
to go to the next entry (or
to go back to the
previous one).
After reaching the entry numbered
22
the display wraps and
returns to the beginning of the error log, displaying
E01
again.
To exit from the error log and return to normal display mode
press
twice.
The
Settings
icon in the menu bar will go out.
Table 7-9 Reading the Error Log
Table 7-10 Entering the Calibration Reference Temperature Manually
Prepare for temperature calibration (see
“Comparison Measurement Procedure”
).
Press
to activate the menu bar, then use
or
to select the
Settings
icon and press
to confirm.
Summary of Contents for Isotemp 178L 120V
Page 1: ......
Page 6: ...Contents iv Peltier Low Temp Incubators Fisher Scientific...
Page 8: ...List of Figures ii Peltier Low Temp Incubators Fisher Scientific...
Page 84: ...Technical Data 13 4 Peltier Low Temp Incubators Fisher Scientific...
Page 88: ...Device Log 15 2 Peltier Low Temp Incubators Fisher Scientific...
Page 91: ...Contact Fisher Scientific Peltier Low Temp Incubators 16 3...