Handling and Control
Settings
Fisher Scientific
Peltier Low Temp. Incubators 7-11
Temperature Display Unit
The
Settings ->°C / °F
menu item allows for toggling the incubator used for displaying
temperatures between degrees Centigrade and Fahrenheit.
Note
This setting does not have any impact on data logging via the RS-232 interface. Any
temperature data that is logged to a computer for operational parameter documentation purposes is
handed over in °C.
Press
to switch to the
CAL(ibration)
menu item.
The word
USEr
appears in the display pane.
Press
to confirm the selection.
In the settings dialog that appears, set the temperature
measured with the external reference sensor by using
or
and confirm your settings with
.
Table 7-10 Entering the Calibration Reference Temperature Manually
Table 7-11 Toggling the Temperature Display Unit
Press
to activate the menu bar, then use
or
to select the
Settings
icon and press
to confirm.
Press
to switch to the
C - F
menu item.
The text
C - F
is flashing in the display pane.