Wave Analyzer Medica 700 User Guide
204
Appendices
Immunity
test
Test level IEC
60601
Compliance
level
Electromagnetic environment
– guidance
Electrical
fast
transient/
burst
IEC 61000-
4-4
± 2 kV for
power supply
lines
± 1 kV for input/
output lines
Complies
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge
IEC 61000-
4-5
± 1 kV
differential mode
±2 kV common
mode
Complies
Mains power quality should be that
of a typical commercial or hospital
environment.
Voltage
dips, short
interruption
s and
voltage
variations
on power
supply input
lines
IEC 61000-
4-11
<5 % UT (>95 %
dip in UT) for 0,5
cycle
40 % UT (60%
dip in UT) for 5
cycles
70 % UT (30 %
dip in UT) for 25
cycles
<5 % UT (>95 %
dip in UT) for 5
sec
Complies
Mains power quality should that of
a typical commercial or hospital
environment. If the user of the
Wave Analyzer Medica 700
requires continued operation
during power mains interruptions,
it is recommended that the Wave
Analyzer Medica 700 be powered
from an uninterruptible power
supply or a battery.
Power
frequency
(50/60
hertz)
magnetic
field
IEC 61000-
4-8
3 A/m
Complies
Power magnetic fields should be
at levels characteristic of a typical
location in a typical commercial or
hospital environment.
Note:
UT is the a.c. mains voltage prior to application of the test level.
Summary of Contents for Medica 700
Page 2: ......
Page 7: ...1 Introduction I...
Page 11: ...2 Safety II...
Page 15: ...3 Equipment and Installation III...
Page 25: ...4 Overview of the Software IV...
Page 38: ...Wave Analyzer Medica 700 User Guide 32 Overview of the Software Ringer Screen...
Page 94: ...5 Managing Patients V...
Page 103: ...6 Performing a Diagnostic VI...
Page 128: ...7 Managing the List of Test Results VII...
Page 135: ...8 Managing the Database VIII...
Page 141: ...9 General Actions and Features IX...
Page 147: ...10 Configuring the Unit X...
Page 189: ...11 What Should I Do If XI...
Page 191: ...12 Maintenance XII...
Page 195: ...189 Wave Analyzer Medica 700 User Guide 189 Maintenance 4 Head and ChinRest are lowered 5...
Page 196: ...Wave Analyzer Medica 700 User Guide 190 Maintenance 6 7...
Page 197: ...191 Wave Analyzer Medica 700 User Guide 191 Maintenance 8 9...
Page 198: ...Wave Analyzer Medica 700 User Guide 192 Maintenance 10 11...
Page 199: ...193 Wave Analyzer Medica 700 User Guide 193 Maintenance 12...
Page 200: ...Wave Analyzer Medica 700 User Guide 194 Maintenance 13...
Page 201: ...195 Wave Analyzer Medica 700 User Guide 195 Maintenance 14...
Page 202: ...Wave Analyzer Medica 700 User Guide 196 Maintenance 15...
Page 203: ...197 Wave Analyzer Medica 700 User Guide 197 Maintenance 16...
Page 204: ...13 Appendices XIII...