Wave Analyzer Medica 700 User Guide
203
Appendices
IEC60601-1 ed.3.1.:
Medical electrical equipment - Part 1: General requirements for
basic safety and essential electrical safety performance.
IEC60601-1-2 ed.3.:
Medical electrical equipment - Part 1-2: General requirements for
safety - Collateral standard: Electromagnetic compatibility – Requirements and tests.
Lifetime of the product: 7 years
1rst CE marking : 2013-12
Electromagnetic Emissions
The Wave Analyzer Medica 700 is intended for use in the electromagnetic environment
specified below. The customer or the user of the Wave Analyzer Medica 700 should assure
that it is used in such an environment.
Emission test
Compliance Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
The Wave Analyzer Medica 700 uses RF energy
only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The Wave Analyzer Medica 700 is suitable for use
in all establishments, including domestic
establishments and those directly connected to the
public low-voltage power supply network that
supplies buildings used for domestic purpose.
Harmonic
emissions CEI
61000-3-2
Class A
Voltage
fluctuations /
flicker emissions
CEI 61000-3-3
Complies
Electromagnetic Immunity
The Wave Analyzer Medica 700 is intended for use in the electromagnetic environment
specified below. The customer or the user of the Wave Analyzer Medica 700 should assure
that it is used in such an environment.
Immunity
test
Test level IEC
60601
Compliance
level
Electromagnetic environment
– guidance
Electrostati
c discharge
(ESD)
IEC 61000-
4-2
±6 kV contact
± 8 kV air
Complies
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.
Summary of Contents for Medica 700
Page 2: ......
Page 7: ...1 Introduction I...
Page 11: ...2 Safety II...
Page 15: ...3 Equipment and Installation III...
Page 25: ...4 Overview of the Software IV...
Page 38: ...Wave Analyzer Medica 700 User Guide 32 Overview of the Software Ringer Screen...
Page 94: ...5 Managing Patients V...
Page 103: ...6 Performing a Diagnostic VI...
Page 128: ...7 Managing the List of Test Results VII...
Page 135: ...8 Managing the Database VIII...
Page 141: ...9 General Actions and Features IX...
Page 147: ...10 Configuring the Unit X...
Page 189: ...11 What Should I Do If XI...
Page 191: ...12 Maintenance XII...
Page 195: ...189 Wave Analyzer Medica 700 User Guide 189 Maintenance 4 Head and ChinRest are lowered 5...
Page 196: ...Wave Analyzer Medica 700 User Guide 190 Maintenance 6 7...
Page 197: ...191 Wave Analyzer Medica 700 User Guide 191 Maintenance 8 9...
Page 198: ...Wave Analyzer Medica 700 User Guide 192 Maintenance 10 11...
Page 199: ...193 Wave Analyzer Medica 700 User Guide 193 Maintenance 12...
Page 200: ...Wave Analyzer Medica 700 User Guide 194 Maintenance 13...
Page 201: ...195 Wave Analyzer Medica 700 User Guide 195 Maintenance 14...
Page 202: ...Wave Analyzer Medica 700 User Guide 196 Maintenance 15...
Page 203: ...197 Wave Analyzer Medica 700 User Guide 197 Maintenance 16...
Page 204: ...13 Appendices XIII...