Performing a Diagnostic
6. 1 Preparing to Run a Diagnostic
, there are a number of things you should do before you
start running the diagnostics:
Clean the headrest. Make sure the pachy filter is clean and use the anti-fog spray on a
weekly basis.
Place a new sheet of protective paper over the chin rest, or clean the chin rest.
In the
Measure
screen, in the list of available measures (#6 in the diagram
),
select the diagnostic you want to perform. The tests needed to perform the selected
diagnostic are highlighted in the list of tests (#18 in the diagram
).
In the upper-right corner (#7 in the diagram
), select whether you want to
perform the diagnostics on the right eye, the left eye, or both.
Under
Position
, select
Adult
if the patient is an adult, or
Child
if the patient is a child.
The unit's head and chin rest move to the default positions for the selected type of
patient.
If you are performing the WF test and/orthe Tonometer test, and you want to modify the
parameters of these tests (the number of measures to take), in the
Measure
screen,
select
Parameters
. In the dialog box that opens, modify the parameters as required, and
then select
OK
to save the changes.
Instruct the patient to sit down, place their chin on the chin rest, and lean their forehead
on the headrest.
If you need to fine-tune the position of the chin rest, select the
Chin Rest
up and down
arrows (#12 in the diagram
) as necessary to move the chin rest up or down.
If you need to adjust the position of the unit's head, select the up, down, right, and left
arrows on the edge of the lens view (#3, #16, #17, and #10 in the diagram
) as
necessary to move the head up, down, right, or left.
Instruct the patient to look at the target (balloon). You should see their eye in the image of
the lens view in the
Measure
screen.
Align the patient's pupil with the center of the target (see
.
Once you have done all of these things, select
Go
to begin the diagnostic.
6. 2 Managing the Diagnostic Procedure
The diagnostic procedure is managed in the
. In this screen, you can select
the diagnostic you want to perform, position the unit head and chin rest so that the patient's eye
is centered in the camera view, modify default test settings, start the diagnostic procedure, and
monitor its progress.
Enhancing the Wavefront Measurement
24
25
25
25
25
25
100
24
100
101
Summary of Contents for Medica 700
Page 2: ......
Page 7: ...1 Introduction I...
Page 11: ...2 Safety II...
Page 15: ...3 Equipment and Installation III...
Page 25: ...4 Overview of the Software IV...
Page 38: ...Wave Analyzer Medica 700 User Guide 32 Overview of the Software Ringer Screen...
Page 94: ...5 Managing Patients V...
Page 103: ...6 Performing a Diagnostic VI...
Page 128: ...7 Managing the List of Test Results VII...
Page 135: ...8 Managing the Database VIII...
Page 141: ...9 General Actions and Features IX...
Page 147: ...10 Configuring the Unit X...
Page 189: ...11 What Should I Do If XI...
Page 191: ...12 Maintenance XII...
Page 195: ...189 Wave Analyzer Medica 700 User Guide 189 Maintenance 4 Head and ChinRest are lowered 5...
Page 196: ...Wave Analyzer Medica 700 User Guide 190 Maintenance 6 7...
Page 197: ...191 Wave Analyzer Medica 700 User Guide 191 Maintenance 8 9...
Page 198: ...Wave Analyzer Medica 700 User Guide 192 Maintenance 10 11...
Page 199: ...193 Wave Analyzer Medica 700 User Guide 193 Maintenance 12...
Page 200: ...Wave Analyzer Medica 700 User Guide 194 Maintenance 13...
Page 201: ...195 Wave Analyzer Medica 700 User Guide 195 Maintenance 14...
Page 202: ...Wave Analyzer Medica 700 User Guide 196 Maintenance 15...
Page 203: ...197 Wave Analyzer Medica 700 User Guide 197 Maintenance 16...
Page 204: ...13 Appendices XIII...