Wave Analyzer Medica 700 User Guide
184
What Should I Do If...?
You may encounter some of the following common problems while working with the Wave
Analyzer Medica 700. If you do, try following the suggested solutions listed below. If the problem
persists and the suggested solution does not remedy it, contact a qualified service
representative or your local distributor.
Problem
Possible Causes
Suggested Solutions
Images are not
properly centered
External lights or
windows near the
unit
Large patches of
light near the unit
Ensure that there no bright lights or
windows near the unit or the patient.
Ensure that there are no large
irregular patches of light on the patient
or on the unit.
Measured pupil size
is different from the
expected or actual
pupil size
External lights or
windows near the
unit
Large patches of
light near the unit
Ensure that there no bright lights or
windows near the unit or the patient.
Ensure that there are no large
irregular patches of light on the patient
or on the unit.
High-order
aberrations that
differ from what is
expected
Direct light falling
on the unit or on
the patient.
Ensure that there no bright lights or
windows near the unit or the patient.
Ensure that there are no large
irregular patches of light on the patient
or on the unit.
The Pachy
measurement
always goes to
manual mode
The pachy filter is
foggy due to the
cold
Use the anti-fog spray at least once a
week, and when the pachy
consistently goes to manual mode.
Tickets do not print
Paper jam
Out of paper
Open the print paper door and release
the paper.
Replace the paper roll.
Summary of Contents for Medica 700
Page 2: ......
Page 7: ...1 Introduction I...
Page 11: ...2 Safety II...
Page 15: ...3 Equipment and Installation III...
Page 25: ...4 Overview of the Software IV...
Page 38: ...Wave Analyzer Medica 700 User Guide 32 Overview of the Software Ringer Screen...
Page 94: ...5 Managing Patients V...
Page 103: ...6 Performing a Diagnostic VI...
Page 128: ...7 Managing the List of Test Results VII...
Page 135: ...8 Managing the Database VIII...
Page 141: ...9 General Actions and Features IX...
Page 147: ...10 Configuring the Unit X...
Page 189: ...11 What Should I Do If XI...
Page 191: ...12 Maintenance XII...
Page 195: ...189 Wave Analyzer Medica 700 User Guide 189 Maintenance 4 Head and ChinRest are lowered 5...
Page 196: ...Wave Analyzer Medica 700 User Guide 190 Maintenance 6 7...
Page 197: ...191 Wave Analyzer Medica 700 User Guide 191 Maintenance 8 9...
Page 198: ...Wave Analyzer Medica 700 User Guide 192 Maintenance 10 11...
Page 199: ...193 Wave Analyzer Medica 700 User Guide 193 Maintenance 12...
Page 200: ...Wave Analyzer Medica 700 User Guide 194 Maintenance 13...
Page 201: ...195 Wave Analyzer Medica 700 User Guide 195 Maintenance 14...
Page 202: ...Wave Analyzer Medica 700 User Guide 196 Maintenance 15...
Page 203: ...197 Wave Analyzer Medica 700 User Guide 197 Maintenance 16...
Page 204: ...13 Appendices XIII...