201
Wave Analyzer Medica 700 User Guide
201
Appendices
Tonometer Specifications
Calibrated Range
7mmHg – 44mmHg
Accuracy
+/-2 mmHg
Method
Non-contact air puff applanation tonometry
Retroillumination Specifications
Measurement wavelength
850nm
Method
Retro led
LCD Screen Specifications
Screen type
TFT LCD
Touchscreen type
Capacitance type
Size
10” WSVGA resolution (1024 × 600 pixels)
Contrast ratio
500
Typical luminosity
250 Cd/m2
In a clinical comparison with a Goldmann tonometer on 231 eyes covering the range 7 –
33 mmHg, 95% of measurements with the Wave Analyzer Medica 700 tonometer were
within ±5mmHg of the Goldmann tonometer measurements.
13. 2 Clinical Studies
Pachymeter Study:
A clinical study was performed including 68 normal eyes that compared the Wave Analyzer
Medica 700 pachymeter function with the Pentacam rotating Scheimpflug camera.
Measurements of central corneal thickness (CCT) were recorded. The mean CCT on the
Wave Analyzer Medica 700 was 541.74 ±39.76µm (std dev) and on the Pentacam was 535.72
± 37.93µm. The Wave Analyzer Medica 700 measures 6.02µm thicker than the Pentacam. A
Bland Altmann graph showing the average and standard deviations of the measurement
differences as well as the 95% confidence levels (std dev × 1.96) and the 99% confidence
levels (std dev × 2.58) is shown in Figure A1. The 95% confidence levels for the measurement
were from 24.54µm to -16.80µm and 99% confidence levels for the measurement were from
Summary of Contents for Medica 700
Page 2: ......
Page 7: ...1 Introduction I...
Page 11: ...2 Safety II...
Page 15: ...3 Equipment and Installation III...
Page 25: ...4 Overview of the Software IV...
Page 38: ...Wave Analyzer Medica 700 User Guide 32 Overview of the Software Ringer Screen...
Page 94: ...5 Managing Patients V...
Page 103: ...6 Performing a Diagnostic VI...
Page 128: ...7 Managing the List of Test Results VII...
Page 135: ...8 Managing the Database VIII...
Page 141: ...9 General Actions and Features IX...
Page 147: ...10 Configuring the Unit X...
Page 189: ...11 What Should I Do If XI...
Page 191: ...12 Maintenance XII...
Page 195: ...189 Wave Analyzer Medica 700 User Guide 189 Maintenance 4 Head and ChinRest are lowered 5...
Page 196: ...Wave Analyzer Medica 700 User Guide 190 Maintenance 6 7...
Page 197: ...191 Wave Analyzer Medica 700 User Guide 191 Maintenance 8 9...
Page 198: ...Wave Analyzer Medica 700 User Guide 192 Maintenance 10 11...
Page 199: ...193 Wave Analyzer Medica 700 User Guide 193 Maintenance 12...
Page 200: ...Wave Analyzer Medica 700 User Guide 194 Maintenance 13...
Page 201: ...195 Wave Analyzer Medica 700 User Guide 195 Maintenance 14...
Page 202: ...Wave Analyzer Medica 700 User Guide 196 Maintenance 15...
Page 203: ...197 Wave Analyzer Medica 700 User Guide 197 Maintenance 16...
Page 204: ...13 Appendices XIII...