Cerabar S/Deltabar S/Deltapilot S 4...20mA HART
Trouble-shooting
Hauser
133
110 (A110)
Alarm
B
Failure (F)
B>Checksum error in
EEPROM: configuration
segment
– The supply voltage is
disconnected when writing.
– Reestablish supply voltage.
Perform reset (Code 7864) if
necessary. Carry out calibration
again.
6
– Electromagnetic effects are
greater than specifications in the
technical data.
See Technical
Information TI00382P (Deltabar
S), TI00383P (Cerabar S) or
TI00416P (Deltapilot S).
– Block off electromagnetic effects
or eliminate sources of
disturbance.
– Main electronics defect.
– Replace main electronics.
113 (A113)
Alarm
B
Failure (F)
B>ROM failure in transmitter
electronic
– Main electronics defect.
– Replace main electronics.
1
115 (E115)
Error
B
factory
setting:
Warning C
Out of
specification
(S)
B>Sensor overpressure
– Overpressure present.
– Reduce pressure until message
disappears.
29
– Sensor defect.
– Replace sensor.
116 (W116)
Warning
C
Maintenance
request (M)
C>Download error, repeat
download
– The file is defect.
– Use another file.
36
– During the download, the data
are not correctly transmitted to
the processor, e.g. because of
open cable connections, spikes
(ripple) on the supply voltage or
electromagnetic effects.
– Check cable connection PC –
transmitter.
– Block off electromagnetic effects
or eliminate sources of
disturbance.
– Perform reset (Code 7864) and
carry out calibration again.
– Repeat download.
120 (E120)
Error
B
factory
setting:
Warning C
Out of
specification
(S)
B>Sensor low pressure
– Pressure too low.
– Increase pressure until message
disappears.
30
– Sensor defect.
– Replace sensor.
121 (A121)
Alarm
B
Failure (F)
B>Checksum error in factory
segment of EEPROM
– Main electronics defect.
– Replace main electronics.
5
122 (A122)
Alarm
B
Failure (F)
B>Sensor not connected
– Cable connection sensor –main
electronics disconnected.
– Check cable connection and
repair if necessary.
13
– Electromagnetic effects are
greater than specifications in the
technical data.
See Technical
Information TI00382P (Deltabar
S), TI00383P (Cerabar S) or
TI00416P (Deltapilot S).
– Block off electromagnetic effects
or eliminate source of
disturbance.
– Main electronics defect.
– Replace main electronics.
– Sensor defect.
– Replace sensor.
130 (A130)
Alarm
B
Failure (F)
B>EEPROM is defect.
– Main electronics defect.
– Replace main electronics.
10
131 (A131)
Alarm
B
Failure (F)
B>Checksum error in
EEPROM: min/max segment
– Main electronics defect.
– Replace main electronics.
9
132 (A132)
Alarm
B
Failure (F)
B>Checksum error in totalizer
EEPROM
– Main electronics defect.
– Replace main electronics.
7
133 (A133)
Alarm
B
Failure (F)
B>Checksum error in History
EEPROM
– An error occurred when writing.
– Perform reset (Code 7864) and
carry out calibration again.
8
– Main electronics defect.
– Replace electronics.
602 (W602)
Warning
C
Funktion
check (C)
C>Linearisation curve not
monotone
– The linearisation table is not
monotonic increasing or
decreasing.
– Add to linearisation table or
perform linearisation again.
57
Code
Error type/
NA 64
Corresponds
NE 107
Message/description
Cause
Measure
Prio
rity