Cerabar S/Deltabar S/Deltapilot S 4...20mA HART
Description of parameters
Hauser
97
TEMP. ENG. UNIT (318)
Selection
Select the unit for the temperature measured values.
See also PCB TEMPERATURE (Page 117) and SENSOR TEMP. (Page 122).
Options:
• °C
• °F
• K
• R
Factory setting:
°C
DENSITY UNIT (001)/(812)
Options
Select density unit.
Options:
• g/cm
3
• kg/dm
3
• kg/m
3
• US lb/in
3
• US lb/ft
3
Factory setting:
kg/dm
3
ADJUST DENSITY
(007)/(316)
Entry
Enter density of fluid.
!
Note!
LIN. MEASURAND: "% (Level)", "Mass" and "Volume" and MEASUAND KOMB.:
If a change to dry calibration is made after a wet calibration using the CALIBRATION
MODE parameter (
Page 73 or 87), the density for this parameter must be entered
correctly before changing the calibration mode. In the event that the pressure falls with
increasing levels (LIN. MEASURED: Volume), such as in the case of a residual volume
measurement, a negative value shall be entered for this parameter.
Factory setting:
1.0
PROCESS DENSITY
(025)/(811)
Entry
Enter a new density value for density correction.
The calibration was carried out with the medium water, for example. Now the container
is to be used for another fluid with another density. The calibration is corrected
appropriately by entering the new density value in the PROCESS DENSITY parameter.
!
Note!
LIN. MEASURAND: "% (Level)", "Mass" and "Volume" and MEASUAND KOMB.:
If a change to dry calibration is made after a wet calibration using the CALIBRATION
MODE parameter (
Page 73 or 87), the density for this parameter must be entered
correctly before changing the calibration mode. In the event that the pressure falls with
increasing levels (LIN. MEASURED: Volume), such as in the case of a residual volume
measurement, a negative value shall be entered for this parameter.
Factory setting:
1.0
Table 16: (GROUP SELECTION
)
OPERATING MENU
SETTINGS
EXTENDED SETUP "Level"
Parameter name
Description