Description of parameters
Cerabar S/Deltabar S/Deltapilot S 4...20mA HART
106
Hauser
LINE-NUMB
Entry
Enter the line number for the linearisation table. A linearisation table must have at least 2
points and may not have more than 32 points.
• TABLE SELECTION = View meas. table
Via this parameter you can select the point of the linearisation table which should be
displayed.
• TABLE SELECTION = Editor table
Enter a point via the LINE-NUMB, X-VAL. and Y-VAL. parameters.
See also this table, parameter description for LIN. EDIT MODE, X-VAL. ("Manual"
entry mode), X-VAL. ("Semiautomatic" entry mode) and Y-VAL.
!
Note!
In the operating program, you can enter a complete linearisation table in one go via the
"Lin.-Tab." window.
X-VAL. ("Manual" entry
mode)
Entry
Enter the pressure value for the linearisation table.
See also LIN. EDIT MODE, LINE-NUMB and Y-VAL.
Prerequisite:
• TABLE SELECTION = Editor table
X-VAL. ("Semiautomatic"
entry mode)
Display
In the "Semiautomatic" entry mode, the container is filled or emptied in stages. The X-
VAL. displays the measured hydrostatic pressure.
Prerequisite:
• TABLE SELECTION = Editor table
Operating program
The X-VAL. is saved by confirming the Y-value.
HART Handheld
Confirm X-VAL. displayed.
See also LIN. EDIT MODE, LINE-NUMB and Y-VAL.
Y-VAL.
Entry
Enter the volume, mass or %-value belonging to the X-VAL. for the linearisation table.
Prerequisite:
• TABLE SELECTION = Editor table
Depending on the setting in the LINd. MEASURAND or COMB. MEASURAND
parameters, enter a volume, mass or %-value here.
See also this table, parameter description for LIN. EDIT MODE, LINE-NUMB, X-VAL.
("Manual" entry mode), X-VAL. ("Semiautomatic" entry mode).
EDITOR TABLE
Options
Select the function for the editor table.
Prerequisite:
• TABLE SELECTION = Editor table
Options:
• Next point: without function
• Last input point: without function
• Accept input table: save editor table as measuring table. This overwrites the old
measuring table.
• Abort: save values entered up to this point for the editor table and display next
parameter. The editor table is not activated as a measuring table.
• Insert point: see example below.
• Delete point: the current point is deleted. See example below.
Example:
Add point, in this case between the 4th and 5th point for example
– Select point 5 via the LINE NUMB parameter.
– Using the TABLE EDITOR parameter, select the option "Insert point".
– Point 5 is displayed for the LINE NUMB parameter. New values for the X-VAL and Y-
VAL parameters.
Example:
delete point, in this case the 5th point for example
– Select point 5 via the LINE NUMB parameter.
– Using the TABLE EDITOR parameter, select the option "Delete point".
– The 5th point is deleted. All of the following points are pushed up one number i.e.
following deletion, the 6th point becomes Point 5.
Factory setting:
Next point
Table 19: (GROUP SELECTION
) OPERATING MENU
SETTINGS
LINEARISATION – Digital
communication
Parameter name
Description