3.9.3
Date
The
Date
command is a user-defined variable that provides a place to save the date of the last
revision of configuration information. It has no impact on the operation of the transmitter or the
HART-based communicator.
3.9.4
Descriptor
The
Descriptor
variable provides a longer user-defined electronic label to assist with more
specific transmitter identification than is available with tag. The descriptor may be up to 16
characters long and has no impact on the operation of the transmitter or the HART-based
communicator.
3.9.5
Message
The
Message
variable provides the most specific user-defined means for identifying individual
transmitters in multi-transmitter environments. It allows for 32 characters of information and is
stored with the other configuration data. The message variable has no impact on the operation
of the transmitter or the HART-based communicator.
3.10
Measurement filtering
3.10.1
50/60 Hz filter
The
50/60 Hz Filter
(also known as Line Voltage Filter or AC Power Filter) command sets the
transmitter electronic filter to reject the frequency of the AC power supply in the plant. The 60
Hz or 50 Hz mode can be chosen. The factory default for this setting is 60 Hz.
Note
In high noise environments, normal mode is recommended.
HART 5 Fast Keys
2, 2, 7, 1, 2
HART 7 Fast Keys
2, 2, 7, 1, 3
HART 5 Fast Keys
2, 2, 7, 1, 3
HART 7 Fast Keys
2, 2, 7, 1, 4
HART 5 Fast Keys
2, 2, 7, 1, 4
HART 7 Fast Keys
2, 2, 7, 1, 5
HART 5 Fast Keys
2, 2, 7, 5, 1
HART 7 Fast Keys
2, 2, 7, 5, 1
Summary of Contents for Rosemount
Page 1: ...Reference Manual 00809 0100 4021 Rev GC April 2014 Rosemount 3144P Temperature Transmitter ...
Page 2: ......
Page 4: ...iv Reference Manual 00809 0100 4021 Rev GC Title Page April 2014 ...
Page 10: ...x Reference Manual 00809 0100 4021 Rev GC Table of Contents April 2014 Table of Contents ...
Page 38: ...28 Reference Manual 00809 0100 4021 Rev GC Section 2 Installation April 2014 Installation ...
Page 188: ...178 Reference Manual 00809 0100 4021 Rev GC Section 5 Maintenance April 2014 Maintenance ...
Page 250: ...Index 244 Reference Manual 00809 0100 4021 Rev GC Index April 2014 Index ...
Page 251: ......