123
Reference Manual
00809-0100-4021, Rev GC
Section 3: HART commissioning
April 2014
Figure 3-11. Trim
3.15.1
Sensor input trim
The
Sensor Trim
command allows for alteration of the transmitter’s interpretation of the input
signal as shown in
. The sensor trim command trims, in engineering (°F,
°C, °R, K) or raw (
, mV) units, the combined sensor and transmitter system to a site standard
using a known temperature source. Sensor trim is suitable for validation procedures or for
applications that require profiling the sensor and transmitter together.
Perform a sensor trim if the transmitter’s digital value for the primary variable does not match
the plant’s standard calibration equipment. The sensor trim function calibrates the sensor to
the transmitter in temperature units or raw units. Unless the site-standard input source is
NIST-traceable, the trim functions will not maintain the NIST-traceability of your system.
The trim functions should not be confused with the rerange functions. Although the rerange
command matches a sensor input to a 4–20 mA output—as in conventional calibration—it does
not affect the transmitter’s interpretation of the input.
Note
A warning will appear “Set the Control Loop to Manual” (see
“Setting the loop to manual” on
.)
Application
: Linear Offset
Application
: Linear Offset and
Slope Correction
Solution
: Single-Point Trim
Solution
: Two-Point Trim
Method
:
1.Connect sensor to transmitter. Place sensor in
bath between range points.
2.Enter known bath temperature using the Field
Communicator.
Method
:
1.Connect sensor to transmitter. Place sensor in
bath at low range point.
2.Enter known bath temperature using the Field
Communicator.
3.Repeat at high range point.
One-Point Trim
Two-Point Trim
Transmitter System Curve
Site-Standard Curve
HART 5 Fast Keys
3, 4, 4
HART 7 Fast Keys
3, 4, 4
R
e
si
st
anc
e
(
o
hm
s)
Temperature
R
e
si
st
anc
e
(
o
hm
s)
Temperature
Summary of Contents for Rosemount
Page 1: ...Reference Manual 00809 0100 4021 Rev GC April 2014 Rosemount 3144P Temperature Transmitter ...
Page 2: ......
Page 4: ...iv Reference Manual 00809 0100 4021 Rev GC Title Page April 2014 ...
Page 10: ...x Reference Manual 00809 0100 4021 Rev GC Table of Contents April 2014 Table of Contents ...
Page 38: ...28 Reference Manual 00809 0100 4021 Rev GC Section 2 Installation April 2014 Installation ...
Page 188: ...178 Reference Manual 00809 0100 4021 Rev GC Section 5 Maintenance April 2014 Maintenance ...
Page 250: ...Index 244 Reference Manual 00809 0100 4021 Rev GC Index April 2014 Index ...
Page 251: ......