148
Reference Manual
00809-0100-4021, Rev GC
Section 4: F
OUNDATION
fieldbus Configuration
April 2014
Foundation™ fieldbus configuration
4.5.3
Recommended actions for PlantWeb alerts
RECOMMENDED_ACTION
The RECOMMENDED_ACTION parameter displays a text string that gives a recommended
course of action based on the type and specific event of the PlantWeb alerts are active.
Table 4-2. RB.RECOMMENDED_ACTION
Alarm type
Failed/maint/advise
active event
Recommended action
text string
P
lan
tW
e
b
ale
rts
None
None
No action required
Advisory
NV Writes Deferred
Non-volatile writes have been deferred, leave the
device powered until the advisory goes away
Maintenance
Configuration Error
Re-write the Sensor Configuration
Primary Value
Degraded
Confirm the operating range of the applied sensor
and/or verify the sensor connection and device
environment
Calibration Error
Retrim the device
Secondary Value
Degraded
Verify the ambient temperature is within operating
limits
Failed
Electronics Failure
Replace the Device
HW / SW Incompatible
Verify the Hardware Revision is compatible with the
Software Revision
NV Memory Failure
Reset the device then download the Device
Configuration.
Primary Value Failure
Verify the instrument process is within the Sensor
range and/or confirm sensor configuration and
wiring.
Secondary Value
Failure
Verify sensor range and/or confirm sensor
configuration and wiring
Diagnostic
Error
Sensor Drift Alert or
Hot BU active
Confirm the operating range of the supplied sensor
and/or verify the sensor connection and device
environment
Primary Value
Degraded
Confirm the operating range of the supplied sensor
and/or verify the sensor connection and device
environment
Summary of Contents for Rosemount
Page 1: ...Reference Manual 00809 0100 4021 Rev GC April 2014 Rosemount 3144P Temperature Transmitter ...
Page 2: ......
Page 4: ...iv Reference Manual 00809 0100 4021 Rev GC Title Page April 2014 ...
Page 10: ...x Reference Manual 00809 0100 4021 Rev GC Table of Contents April 2014 Table of Contents ...
Page 38: ...28 Reference Manual 00809 0100 4021 Rev GC Section 2 Installation April 2014 Installation ...
Page 188: ...178 Reference Manual 00809 0100 4021 Rev GC Section 5 Maintenance April 2014 Maintenance ...
Page 250: ...Index 244 Reference Manual 00809 0100 4021 Rev GC Index April 2014 Index ...
Page 251: ......