161
Reference Manual
00809-0100-4021, Rev GC
Section 4: F
OUNDATION
fieldbus Configuration
April 2014
Foundation™ fieldbus configuration
4.9.7
Analog input diagnostics
Table 4-9. AI BLOCK_ERR Conditions
Table 4-10. Troubleshooting the AI block
Condition
number
Condition name and description
0
Other
1
Block Configuration Error: the selected channel carries a measurement
that is incompatible with the engineering units selected in XD_SCALE,
the L_TYPE parameter is not configured, or CHANNEL = zero.
3
Simulate Active: Simulation is enabled and the block is using a
simulated value in its execution.
7
Input Failure/Process Variable has Bad Status: The hardware is bad, or a
bad status is being simulated.
14
Power Up: Block is not scheduled.
15
Out of Service: The actual mode is out of service.
Symptom
Possible causes
Recommended actions
Bad or no temperature
readings
(Read the AI “BLOCK_ERR”
parameter)
BLOCK_ERR reads OUT OF
SERVICE (OOS)
1. AI Block target mode target mode set to OOS.
2. Resource Block OUT OF SERVICE.
BLOCK_ERR reads
CONFIGURATION ERROR
1. Check CHANNEL parameter (see
2. Check L_TYPE parameter (see
)
3. Check XD_SCALE engineering units. (see
)
BLOCK_ERR reads POWERUP
Download schedule into block. Refer to host for
downloading procedure.
BLOCK_ERR reads BAD INPUT
1. Sensor Transducer Block Out Of Service (OOS)
2. Resource Block Out of Service (OOS)
No BLOCK_ERR but readings
are not correct. If using Indirect
mode, scaling could be wrong.
1. Check XD_SCALE parameter.
2. Check OUT_SCALE parameter.
(see
“XD_SCALE and OUT_SCALE” on page 156
No BLOCK_ERR. Sensor needs
to be calibrated or Zero
trimmed.
See
Section 3: HART® commissioning
to determine the
appropriate trimming or calibration procedure.
OUT parameter status reads
UNCERTAIN and substatus
reads EngUnitRangViolation.
Out_ScaleEU_0 and EU_100
settings are incorrect.
See
Summary of Contents for Rosemount
Page 1: ...Reference Manual 00809 0100 4021 Rev GC April 2014 Rosemount 3144P Temperature Transmitter ...
Page 2: ......
Page 4: ...iv Reference Manual 00809 0100 4021 Rev GC Title Page April 2014 ...
Page 10: ...x Reference Manual 00809 0100 4021 Rev GC Table of Contents April 2014 Table of Contents ...
Page 38: ...28 Reference Manual 00809 0100 4021 Rev GC Section 2 Installation April 2014 Installation ...
Page 188: ...178 Reference Manual 00809 0100 4021 Rev GC Section 5 Maintenance April 2014 Maintenance ...
Page 250: ...Index 244 Reference Manual 00809 0100 4021 Rev GC Index April 2014 Index ...
Page 251: ......