Commissioning
The 3144P must be configured for certain basic variables to operate. In many cases, these
variables are pre-configured at the factory. Configuration may be required if the variables need
to be changed.
Commissioning consists of testing the transmitter and verifying transmitter configuration data.
Rosemount
®
3144P transmitters can be commissioned either before or after installation.
Commissioning the transmitter on the bench before installation using a field communicator or
AMS
™
Device Manager ensures that all transmitter components are in working order.
For more information on using the field communicator with the 3144P transmitter, see
. For more information on using the 3144 with F
OUNDATION
™
fieldbus, see
Section 4: Foundation™ fieldbus configuration
Figure 2-1. Installation Flowchart
2.2.1
Setting the loop to manual
Set the process application loop to manual when sending or requesting data that would disrupt
the loop or change the output of the transmitter. The field communicator or AMS Device
Manager will prompt to set the loop to manual, when necessary. Acknowledging the prompt
does not set the loop to manual, it is only a reminder. Setting the loop to manual is a separate
operation.
No
Yes
No
Yes
Bench
Calibration?
BASIC SETUP
Refer to
Set Units
Set Range
Values - HART
Set Sensor
Types
Set Number of
Wires
VERIFY
Simulate Sensor
Input
FIELD INSTALL
Set Jumpers or
Switches
Mount the
Transmitter
Power the
Transmitter
Check for
Process Leaks
Wire the
Transmitter
START HERE
Done
Within
Specifications?
Set Damping
Summary of Contents for Rosemount
Page 1: ...Reference Manual 00809 0100 4021 Rev GC April 2014 Rosemount 3144P Temperature Transmitter ...
Page 2: ......
Page 4: ...iv Reference Manual 00809 0100 4021 Rev GC Title Page April 2014 ...
Page 10: ...x Reference Manual 00809 0100 4021 Rev GC Table of Contents April 2014 Table of Contents ...
Page 38: ...28 Reference Manual 00809 0100 4021 Rev GC Section 2 Installation April 2014 Installation ...
Page 188: ...178 Reference Manual 00809 0100 4021 Rev GC Section 5 Maintenance April 2014 Maintenance ...
Page 250: ...Index 244 Reference Manual 00809 0100 4021 Rev GC Index April 2014 Index ...
Page 251: ......