been inactive for one minute, the output reverts to the process gas. When you have
initiated a calibration, the value at TP5 and TP6 is the % O
2
seen by the cell.
Oxygen levels, as seen on the multimeter, are:
•
8.0% O
2
= 8.0 Vdc
•
0.4% O
2
= 0.4 Vdc
5.2.5
Cal
The
Cal
key can:
•
Initiate a calibration.
•
Sequence through calibration.
•
Abort the calibration.
Keys
Refer to the
Remote Powered Loop LCD
Reference Manual for calibration and operation.
Start up and operate transmitter with membrane keypad
Reference Manual
October 2019
00809-0100-4340
90
Emerson.com/Rosemount