2.5
Rosemount
™
IMPS 4000 connections
Ensure that the Rosemount IMPS 4000 is installed in a safe (non-hazardous and explosive-
free) area and verify the wiring and pneumatic connections per the Rosemount IMPS 4000
Intelligent Multi-probe Test Gas Sequencer Reference Manual.
2.6
Rosemount SPS 4001B connections
Ensure that the Rosemount SPS 4001B is installed in a safe (non-hazardous, explosive-free)
area and verify the wiring and pneumatic connections per the Rosemount SPS 4001B
Single Probe Autocalibration Sequencer Reference Manual.
CAUTION
Upon completing installation, ensure that the Rosemount Oxymitter 4000 is turned
on and operating prior to firing up the combustion process.
Damage can result from having a cold Rosemount Oxymitter 4000 exposed to the process
gases.
During outages, if possible, leave all Rosemount Oxymitter 4000 units running to prevent
condensation and premature aging from thermal cycling.
WARNING
If the ducts will be washed down during outage, ensure that the transmitter units are
powered down and removed from the was area.
Install
Reference Manual
October 2019
00809-0100-4340
68
Emerson.com/Rosemount