Figure 8-21: O2 CELL LED on Membrane Keypad
Potential cause
The EEprom has been changed for a later version and at power-up, the EEprom does not
update.
Recommended actions
1. Power down.
2. Restore power.
Potential cause
If the alarm occurs while the unit is running, there is a hardware problem on the
microprocessor board.
Recommended action
Replace the electronic assembly.
See
Fault 12, LOI
When Fault 12 is detected, the LOI displays the
EEprom Corrupt
message.
Troubleshoot
Reference Manual
October 2019
00809-0100-4340
138
Emerson.com/Rosemount